Comparison of Aortic Annulus Dimensions by Multimodal Measurement Before Transcatheter Aortic Valve Replacement

Ruixiang Qi,Xiangdong You
DOI: https://doi.org/10.1016/j.ultrasmedbio.2017.08.1170
IF: 3.694
2017-01-01
Ultrasound in Medicine & Biology
Abstract:To compare aortic annular diameter measured by transthoracic echocardiography (TTE), transesophageal echocardiography (TEE), and multi-slice spiral computed tomography (MSCT) in patients with severe aortic stenosis, and to evaluate the impact on selection of prosthetic valve size in transcatheter aortic valve replacement (TAVR). Clinical data of 138 patients with severe aortic stenosis referred for TAVR between January 2014 and June 2016 in our hospital were retrospectively analyzed. The difference of aortic annular diameter measured by TTE, TEE, and MSCT were compared. TTE was performed after TAVI to evaluate the accuracy of measurement before TAVI. (1) Mean aortic annulus diameter was 23.37±2.22 mm using TTE and 23.52±1.70 mm using TEE (p = 0.12) and correlation between TTE and TEE was excellent (r = 0.87, p < 0.05). (2) The long-axis diameter (27.86±2.87 mm) and the short diameter (21.91±2.53 mm) measured according to MSCT multiplanar reconstruction were significantly different (p < 0.05). There was no significant difference between the mean of the long- and short-axis diameters (24.92±2.38 mm) and the diameter derived from cross-sectional area (24.84±2.25 mm) by MSCT (p = 0.08). However, the diameter derived from the circumference (25.35±2.34 mm) was significantly larger than the mean of the long- and short-axis diameters and the diameter derived from cross-sectional area by MSCT, and the difference were 0.43±0.62 mm and 0.51±0.62 mm respectively(both p < 0.05).(3) Measurements of the aortic annulus diameter by MSCT including the mean of the long- and short-axis diameters, diameter derived from cross-sectional area, and diameter derived from the circumference were larger than the measurement by TTE and TTE significantly (all p < 0.05).(4) TAVR postoperative follow-up: Implantation was successful in all patients. Moderate to severe paravalvular leakage was detected in 3 patients at 7 days post TAVR, and 1 patient developed severe prosthetic valve restenosis at 6 months post TAVR and received valve-in-valve implantation. In severe aortic patients referred for TAVR, measurements of the aortic annulus diameter using TTE and TEE are smaller than the measurements by MSCT. In the absence of a gold standard, selection of prosthetic valve type in TAVI procedure should rely on comprehensive considerations, which is of importance to get good clinical results for severe aortic stenosis patients underwent TAVI.
What problem does this paper attempt to address?