A Discussion on the Affecting Factors of the Fitting Procedures' Reliability in Isothermal Titration Calorimetry Analysis.

Junjie Chen,Wenjing Tian,Ye Yun,Yingpu Tian,Cuiling Sun,Rong Ding,Haifeng Chen
DOI: https://doi.org/10.1016/j.abb.2021.109045
IF: 4.114
2021-01-01
Archives of Biochemistry and Biophysics
Abstract:The standard isothermal titration calorimetry (ITC) curve, characterized as a typical sigmoid is strictly confined by the so-called c value, which is a ratio of titrand concentration to K-D. The proper c value with a range from 5 to 500 is commonly recommended as a standard protocol in routine detection process for acquiring the reliable fitting results in 1:1 binding mode. However, if the c value is less than "1" due to the weak binding or low concentration of analyte, fitting precision gets unstable and susceptible to the data noise. Herein, we first got a deep discussion into the reliability of the fitting procedure for 1:1 binding mode by data simulation, then quantized the effect of several affecting factors on the precision of parameters estimation through mathematical analysis. Finally, we proposed the value of 2 similar to 4 times K-D for final ligand concentration is optimal for the ITC titration in low c system (c < 1). All the theoretical derivations were further verified by a practical experiment of Magnesium-EDTA binding test.