A Dynamic Measurement System for 3-D Upper-limb Forces During Paraplegic FES-assisted Walking

MING Dong,WAN Bai-kun,HU Yong,LIANG Zhi-ren
DOI: https://doi.org/10.3969/j.issn.1001-4888.2005.01.014
2005-01-01
Abstract:Functional electrical stimulation (FES) was commonly prescribed to resume paraplegic patient walking ability. Up to now, as an important index for paraplegic walking efficiency evaluation, the research on 3-D upper-limb forces was limited and starved for development. This paper suggested a new dynamic measurement system for 3-D upper-limb forces during FES-assisted paraplegic walking, in which a standard walker based on 12 string-gauge bridges was tethered to a computer through a dedicated isolation amplifier and A-D converter. Then after calibration, upper limb forces data could be real-time monitored and displayed. The dynamometer system adopted some special designs, e.g. string-gauge allocation and redundant-optimized method, to guarantee the reliability and precision of measurements. In this design, force accuracy is higher than 1.01%, nonlinearity is less than 2.901%, and crosstalk is less than 3.188%. A pilot trial from a paraplegic volunteer undergoing normal walking training protocol with FES was given in this study to demonstrate the potential clinical usefulness of this method. The results show that this system may be used as 1) an evaluation index of FES-assisted paraplegic walking efficiency, 2) a control indicator during FES-assisted paraplegic walking training and 3) a feed-back signal to choose an efficient FES pattern and sequence.
What problem does this paper attempt to address?