Abstract:Based on luminol-H2O2-Co2+ chemiluminescence system, a new microflow injection chip was designed for the determination of hydrogen peroxide in rainwater. The chip was produced by using two 50 mm x 40 mm x 5 mm in dimension transparent poly (methylmethacrylate) (PMMA) and the channels were etched. by CO2 laser in 200 mu m width and 150 mu m depth. The linear range of hydrogen peroxide concentration was 0.05 similar to 2 mu mol/L and the detection limit was 0.02 mu mol/L (S/N = 3). The relative standard deviation was 4.6% for 0.6 mu mol/L of hydrogen peroxide (n = 9). The proposed method has been successfully applied to the determination of hydrogen peroxide in rainwater.
What problem does this paper attempt to address?