Platinum-Tetra (4-Bromophenyl) Porphyrin Polyvinyl Chloride Membrane Optical Dissolved Oxygen Sensor

L Zhao,Xl Lu,Hy Yuan,Ky Wong,D Xiao
DOI: https://doi.org/10.3321/j.issn:0253-3820.2004.06.004
IF: 1.193
2004-01-01
Chinese Journal of Analytical Chemistry
Abstract:This paper describes the preparation of platinum-tetra (4-bromophenyl) porphyrin (PtTBPP) and the sensing characteristics of the dissolved oxygen photosensor with PtTBPP as sensing element and PVC as support system. The membrane works on the molecular oxygen quenching the fluorescence of the metalloporphyrin at the wavelength of 664 run when excited at the wavelength of 402 rim. It shows a very high sensitivity to dissolved oxygen with the ratio of I-0/I = 15 where I-0 is the fluorescence intensity in 5% Na2SO3 solution (deoxygenated water) and I is the fluorescence intensity in air saturated water. The membrane has good reversibility. Its response to repeat cycles of air saturated water and nitrogen saturated water is complete reversible. During the oxygen concentration range of 0 similar to 0.17 mL/m(3), the membrane's fluorescence intensity decreased by about 11% which demonstrates that it is very sensitive to trace oxygen in water. The photostability of the membrane is extremely good. There is no sign of photodegration when it is exposed to xeon lamp for 1 h in 5% Na2SO3 solution. The membrane has also been used to the analysis of water sample, its accuracy is satistactory compase with the Winkler method.
What problem does this paper attempt to address?