Development and Application of Dissolved Oxygen Analyzer Based on Fluorescence Quenching Principle

LI Wen-long,JING Miao,WANG Zheng,WANG Xiao-ru
DOI: https://doi.org/10.3969/j.issn.1671-6647.2007.03.014
2007-01-01
Abstract:A sensor membrane sensitive to dissolved oxygen has been developed,and the main fluorescent material is an organic Ru complex.When excited by a strong light with a central wavelength of 465 nm,the sensor membrane can generate a fluorescence with a central wavelength of 620 nm,the dissolved oxygen in water can quench the excited fluorescence,and the quenching degree has a linear relationship with the dissolved oxygen concentration.The fluorescence strength of sensor membrane in the nitrogen-saturated water is about 20 times that in the oxygen-saturated water,so making a comparison between the fluorescence signal strengths in a water sample before and after quenching will determine the dissolved oxygen concentration in the water sample.Based on the principle and the fluorimetric technology,an optical dissolved oxygen analyzer was developed and used to test 6 different natural water samples,and the results determined with the analyzer have no significant difference from those with the standard iodimetric method with an error of no greater than 2% and a response time of less than 1 minute.It is shown from the 36-hour-underwater stability test results that the mean deviation of the measued dissolved oxygen values is less than 0.2 mg/L,which indicates that the analyzer is applicable to adverse environment and has a better stability.