Simultaneous Determination of Arsenic,antimony,selenium and Mercury in Water Samples by Chemical Vapor Generation-Four-channel Non-Dispersive Atomic Fluorescence Spectrometry

李海涛,李中玺,杨晓明,郭跃安,张莹,冯玉怀
DOI: https://doi.org/10.3969/j.issn.1000-0720.2009.04.003
2009-01-01
Abstract:In this work,a robust analytical procedure for the simultaneous determination of arsenic,antimony,selenium and mercury in water samples using chemical vapor generation-four-channel non-dispersive atomic fluorescence spectrometry has been developed.The conditions of instrumentation and vapor generation of arsenic,antimony,selenium and mercury were optimized.Hexavalent Se was reduced to tetravalent Se by heating the sample with HCl,and then pentavalent As and Sb were reduced to the trivalent state by the addition of 5 g/L thiourea.Under optimal conditions,the detection limits for As,Sb,Se and Hg were determined to be 0.05,0.03,0.05 and 0.01 ng/mL,respectively.The precisions for 7 replicate determinations at the 5 ng/mL of As,Sb,Se and 0.5 ng/mL of Hg were 0.42%,0.74%,0.97% and 1.00%(RSD),respectively.The proposed method was successfully applied to the simultaneous determination of As,Sb,Se and Hg in various water samples and compared with the standard methods, and spike recoveries were also performed.All the results of the present method were statistically equivalent to the results of the standard method,and the spike recoveries were in the range of 93% to 105%.
What problem does this paper attempt to address?