Simultaneous Arsenic and Selenium Speciation Analysis Using HPLC-ICP-MS

Xie Zong-liang
Abstract:Four arsenic species and three selenium species simultaneous analysis method was build with High performance liquid chromatography and Inductively coupled plasma-mass spectrometry(HPLC-ICP-MS).The column of Hamilton PRP-X100 was used for chromatogram seperation,and Collision cell technology(CCT) was used for ICP-MS analysis.The detection limits for arsenite(As(Ⅲ)),dimethylarsinic acid(DMA),monomethylarsonic acid(MMA),arsenate(As(Ⅴ)),selenocystine(SeCys),selenite(Se(Ⅳ)) and selenomethionine(SeMet) were 0.2 μg/L,0.2 μg/L,0.2 μg/L,0.3 μg/L,0.5 μg/L,0.4 μg/L and 2.5 μg/L,respectively.The line range was 10~200 μg/L for all species except the SeMet,all linear correlation coefficients were not less than 0.9995.The species of arsenic and selenium in fresh mushroom and fish was detected with this method,good spiked recoveries from 81.6% to 103.7% were obtained.
Chemistry,Environmental Science
What problem does this paper attempt to address?