Study of Inhibited Chemiluminescence Reaction by Hydroquinone on Microchip

XU Yi,LAN Yu-wei,ZHANG Xiao-feng,LU Hao
DOI: https://doi.org/10.3969/j.issn.1000-0720.2005.04.009
2005-01-01
Abstract:The miniaturized total analytical system (μTAS) which indirectly detected the concentration of hydroquinone was established by means of flow injection analysis (FIA) technique and the inhibited reaction of hydroquinone to Luminol-H_2O_2-Co~(2+) chemiluminescence on designed PMMA microchip. Different factors affecting the inhibited CL reaction, especially concentration of hydroquinone, were optimized. In order to design a microchip suitable to this reaction system, the design of chip configuration and the process of these reactions were studied. The chemiluminescence intensity had a linear relation ship with the concentration of hydroquinone in water samples, and the detection limit of this method was 2.0×10~(-9) mol/L and the linear range was between 2.0×10~(-7) and 2.0×10~(-8) mol/L.