Miniaturized Point Discharge-Atomic Emission Spectrometry for Detection of Microamount of Iodine

Hong LUO,An-qin LEN,Cheng-bin ZHENG,Xian-deng HOU
DOI: https://doi.org/10.13526/j.issn.1006-6144.2017.05.014
2017-01-01
Abstract:A miniaturized atomic emission spectrometer composed of a point discharge source and a charge-coupled device(CCD) spectrometric detector was constructed and used for the sensitive detection of microamount of iodine.By using hydrogen peroxide,iodide ions were oxidated to iodine vapor,and then the chemical vapor was introduced into the point discharge area with the carrier gas flow of argon.The optical signals were captured and analyzed by the CCD at 206 nm.Under the optimized conditions including 2% hydrogen peroxide,discharge voltage of 90 V,and carrier gas flow rate of 450 mL·min-1,the stable detection signals were obtained(RSD=2.5 %).The calibration curves were plotted with the concentration of iodine ranging from 0.5 to 500 mg·L-1,with a correlation coefficient >0.99 and a limit of detection of 0.1 mg· L-1.The miniaturized instrumentation demonstrated several advantages including simple device,low power consumption,high sensitivity,and portability for field analysis.
What problem does this paper attempt to address?