Rapid detection of aromatic compounds in ambient air by multi-capillary column-ion mobility spectrometry

Zanzan DU,Xiaoguang GAO,Jian JIA,Xiuli HE,Jianping LI
DOI: https://doi.org/10.3969/j.issn.1008-6145.2018.03.003
2018-01-01
Abstract:Multi-capillary column (MCC)-ion mobility spectrometer (IMS) was used for rapid detection of benzene, toluene, ethylbenzene, xylene and styrene in ambient air. The temperature of IMS drift tube was 80℃, and the carrier gas and drift gas were zero air. The reduced mobilities of benzene, toluene, ethylbenzene, xylene and styrene were measured, and the regression equations were acquired in the range of 0.20-3.00, 0.20-2.00, 0.15-1.00, 0.25-3.00 and 0.08-1.00 mg/m3 with the corelation coefficients not more than 0.96. The detection limits based on 3 times the baseline noise were 0.055, 0.023, 0.017, 0.065 and 0.013 mg/m3, respectively. The mixture of benzene, toluene, ethylbenzene, xylene and styrene was analyzed by MCC-IMS, and the relative errors of benzene, toluene and styrene were 5.0%,9.4% and 0.2% with the relative standard deviation of 2.7%, 5.0% and 6.7%(n=6). Each detection could be completed within three minutes.