Study on bisphenol Aelectrochemical aptasensor based on Au-MoS2 and hemin

Meng-fan JIANG,Tian-fang KANG,Li-ping LU
DOI: https://doi.org/10.3969/j.issn.1004-1656.2019.02.013
2019-01-01
Abstract:In this paper, an electrochemical aptamer sensor for bisphenol A (BPA) was fabricated based on the competition bindingof BPA and its aptamer complementary strand (cDNA) to the aptamers.A composite composed of gold nanoparticles and molybdenum disulfide was prepared and was denoted as Au-MoS2.The composite was then modified onto the surface of a glassy carbon electrode (GCE).The thiol-modified cDNA was hybridized with the aptamer to form double-stranded DNA (dsDNA).The dsDNA was modified onto the surface of the electrode based on Au-S bonds produced by chemisorptions of thiol-modified cDNA on Au nanoparticles.Hemin which possesses the porphyrin plane structure was embedded into the dsDNA groove to prepare the electrochemical aptasensor (Hemin-dsDNA/Au-MoS2/GCE).An analytical method for BPA was established by using differential pulse voltammetry (DPV).The reaction between H2O2 and hydroquinone (HQ) was electrochemically catalyzed by hemin.Because the affinity of BPA with aptamers is stronger, under certain experimental conditions, the higher concentration of BPA in solutioncan cause more hemin to fall off the surface of the electrode.The results showed that the peak current value of DPV in the range of 1.0 nmol·L-1 to 10.0 μmol·L-1 BPA was linearly decreased with the increase of BPA concentration under the optimized experimental conditions.The detection limit was 0.8 nmol·L-1.The stability and selectivity of the sensor were investigated.BPA in real sampleswasdetermined, and satisfactory results were obtained.
What problem does this paper attempt to address?