Effects of Hepatitis B Virus Target Concentration on New Style Peptide Nucleic Acid Piezoelectric Gene Sensor Array

Ming CHEN,Wei-ling FU,Guo-ru CAI,Li-li YU,Ming-hua Liu,Qing-hai CHEN,Bo ZHANG,Tian-lun JIANG,Rong WU,Feng WANG
DOI: https://doi.org/10.3321/j.issn:1005-4529.2005.04.005
2005-01-01
Abstract:OBJECTIVE To investigate the effects of hepatitis B virus (HBV) target concentration on hybridization efficiency of peptide nucleic acid (PNA) piezoelectric gene sensor array. METHODS Bis-PNA probe of (1.5 μmol/L) was firstly immobilized on the surface of piezoelectric gene sensor array. Then target HBV genomic DNA varied from 1 pg/L to 100 μg/L was hybridized with the probe, and the decrease in frequency and equilibrium time of reaction were recorded and analyzed. Linear regression was done between frequency decrease and the target concentration; hence the linear range was determined. RESULTS Any frequency change could not be detected by PNA biosensor when target concentration was 1 pg/L. It enhanced firstly and then tended gently when target concentration varied from 10pg/L to 100μg/L, and 10μg/L was the watershed. And no evident change tendency was observed on equilibrium time for hybridization. The statistic linear regression equation was lgC=-2.7455+0.0691×△F among the range of target concentration from 10pg/L to 10μg/L and the correlating coefficient was 0.9923. CONCLUSIONS The frequency change caused by hybridization reaction is a typical saturation curve in accordance with the increase in target concentration. And the linear range of the target concentration is between 10 pg/L to 10μg/L.
What problem does this paper attempt to address?