Research on influence of intravenous contrast agent on dose calculations of helical tomotherapy plans for nasopharyngeal carcinoma

解传滨,徐寿平,戴相昆,王小深,丛小虎,巩汉顺,葛瑞刚,鞠忠建
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2016.01.026
2016-01-01
Abstract:Objective:To investigate the influence of computed tomography(CT) values increased by intravenous contrast agent(CA) on dose calculations of Helical Tomotherapy(HT) plans for nasopharyngeal carcinoma(NPC).Methods: Eight patients with NPC treated by HT were enrolled. Eah patient took two sets of CT in the same position before and after intravenous CA injections. Dose recalculations were done using the C+ images which were uses as a substitute for the C- images of original plans in the QA mode. The 10 points were selected randomly from every target and organ of interest, the CT values and the point doses of the QA plans were compared with the original plans in the same coordinate system.Results: The dose deviation of targets in C- and C+ plans was 0.31%, The vessels indicated the highest difference(1.13%), the dose difference of parotids was 0.55%, the difference of brainstem was the smallest with 0.39%.Conclution: The influence of intravenous contrast agent on dose calculations of HT plans for NPC is small. The C+ images can be used in HT planning theoretically. But the CT values of some tissues such as vessels were increased significantly, which may lead to the errors on the automatic registration of MVCT before delivery. So it needs to be careful in clinical use.