Abstract:The increase in pulsatility of centrifugal pumps can decrease the resistance of blood circulation. Most centrifugal pumps in use now work with constant pressure, which provides nonpulsatile perfusion. In non-pulsatile pumps, the pulsatility of the aortic pressure decreases with the increase of rational speed. The low pulsatile value of aortic pressure has a negative effect on the structures and functions of blood vessels. This paper builds the cardiovascular-pump model and simulates the model using MATLAB. The fuzzy PI Controller, which uses the volatility of the speed of pumps to enhance the pulsatility of the aortic pressure, was designed. Then we set up the experiment platform and linked the blood pump and the controller based on STM32 to the mock cardiovascular system. The result of the experiment indicates that the pulse pressure is 20 mmHg and the mean pressure is 100 mmHg, which can meet the hemodynamic requirements of LVAD.
What problem does this paper attempt to address?