Electrical Safety Testing for the Infant Incubator

Yun-yun WU,Yun-ming SHEN,Chang SU,Qi WANG,Tai-shi ZHENG
DOI: https://doi.org/10.3969/j.issn.1674-1633.2016.09.036
2016-01-01
Abstract:Regular electrical safety testing is an essential part in quality control of all in-use medical equipment. In this paper, 50 infant incubators were tested by using ESA620 automatic electrical safety analyzer based on the standard of IEC60601-1, including grounding resistance protection, insulation resistance, earth leakage current, shell leakage current and other testing items. It could be seen from the testing results that only the grounding resistance protection met the standard (the up-to-standard rate was 70.0%). The main potential safety problems of the infant incubators were the deteriorated condition of the power cords and the oxidization of the outermost shell. According to the post-maintenance test results, the pass rate was 100%.