Report of National Proficiency Testing for Urinary Iodine among the Laboratories of Iodine Deficiency Disorders in 2009

李秀维,李淑华,王海燕,刘列钧,谷云有,王建强,李素梅
2010-01-01
Abstract:Objective The proficiency testing was designed to underst an d the sustained performance ability of urinary iodine labs at different levels a nd the operating situation of quality control network for iodine testing.M ethods The testing results coming from the participatory labs were evalua ted by using standard Z score generated from all the participatory labs.Re sults All of the 32 provincial participatory labs had fed back their test ing results,and 31of them were qualified.At the prefecture level,the feedback rate was 98.6%,and qualified rate was 96.7%.Tianjin Medical University and Ce nter for Endemic Disease Control China CDC got through the proficiency testing.Conclusions The results of the provincial labs are quite close,and the variations inside each lab and inter-lab are small.In addition,the accurac y and precision are maintained at a high level.The testing ability of most pref ecture labs has been significantly improved.At county level,72 labs participat ed in the network and 71 labs passed the proficiency testing.It is suggested th at internal quality control should be emphasized and reference materials should be routinely used in labs at different levels.
What problem does this paper attempt to address?