The Traction of the Modular Tractive Type Splint on Distal Radius Fractures:A Biomechanical Testing and Evaluation

Bo KONG,Yuntao HE,Wei YAN,Xiaobing XI
2018-01-01
Abstract:Objective:To investigate the biomechanical changes of the modular tractive type splint and evaluate the effects of traction.Methods:1 upper extremity specimen was made model of type A2 fracture of the distal radius,and fixed with the modular tractive type splint after a manual reduction maneuver;the small splint has five gears,the total length is 20 mm.The kirschner wire with label was pinned into the distal end fracture.Laser displacement sensor and force sensor were applied to study the effects of traction and the relation of traction with time.Results:With the increasing of length of the traction of the small splint,kirschner wire moved to the distal end;the tractive length of kirschner wire was 9 mm totally.Conclusion:The modular tractive splint has effective traction and keeps the traction effectively.
What problem does this paper attempt to address?