Horizontal belt plate fixation for treating the isolated posterolateral tibial plateau fracture: a biomechanical study

Sun-jun HU,Shi-min ZHANG,Shuang LI,Shou-chao DU,Zhuo MA,Wen-feng XIONG,Ying-qi ZHANG
DOI: https://doi.org/10.13418/j.issn.1001-165x.2018.01.017
2018-01-01
Abstract:Objective To compare and analyze the biomechanical stability of 3 different fixations forthe posterolateral tibial plateau fracture.Methods The posterolateral shearing fracture models were created in 18 synthetic tibia(Synbone) and randomly assigned into 3 groups.The fractures were fixed by horizontal belt plate (Group A),lateral low-profile anatomical LCP (Group B) and posterior reconstructive plate(Group C).Vertical displacements of the posterolateral fragments and failure loads were measured under axial loads including static loading test (from 300 N to 1050 N),fatigue loading test and damage loading teat.Results The vertical displacements in groups A,B,and C were respectively (2.77± 1.79)mm,(2.69± 1.14)mm,and (1.62±0.60)mm under fatigue loading test,the total load 1050N with 10 mm/min and repeat load 10000 times.There was no significant difference between the groups (P>0.05).After the damage loading test,the fmal force in groups A,B,and C were respectively (2055±263) N,(1968±209) N and (2272±130) N,and there was significant difference among the groups (P<0.05).Conclusion Horizontal belt plate fixation for isolated posterolateral tibial plateau fracture can bring sufficient stability for the posterolateral fragment.