Common Influencing Factors in Measurement of Glomerular Filtration Rate Using ~(99M)tc- DTPA Dynamic Imaging

WU Ha,SHI Hong-cheng,LIU Wen-guan,GU Yu-can,ZHU Wei-min,YU Hao-jun,CHEN Shu-guang
DOI: https://doi.org/10.3760/cma.j.issn.1674-1927.2009.06.018
2009-01-01
Abstract:Objective To explore the possible factors influencing the accuracy of glomerular filatration rate (GFR) measurement by ~(99m)Tc-DTPA dynamic renal imaging.Methods Three hundred and twenty six patients and one healthy volunteer underwent ~(99m)c-DTPA dynamic renal imaging to measure GFR.The patients received 500 ml fluid at 30 to 60 minutes before the examination.For the volunteer, 500 ml fluid was given at 5 minutes before the first examination and at 30 minutes before the second examination.The GFRs were calculated using Gates method.The final results were analyzed combined with renogram and GFR.Results Errors in GFR measurement due to unmatched doses of radioactive contrast used and detected were indicated in 61 patients (18.7%), among which, radioactive leakage at the sites of injection was found in 88.5% (54/61) and radioactive retention distally to a tight sleeve was found in 8.2% (5/61). The radioactivity counts of syringe cover were ignored in 2 patients (3.3%), leading to lowered GFR readings.Injection of radioactive agent shortly after fluid-drinking was associated with lowered maximal counts of the curves, elevated excretory phase of the renogram, and thus reduced GFR.The GFR increased abnormally in 65 patients with diabetes, but the renograms curves were normal.Conclusions GFR measurement may be affected by a number of factors.Comprehensive analysis of GFR and renograms is important to ensure the accuracy of tests.Adding patient's history and kidney imaging study would further improve the diagnostic accuracy.
What problem does this paper attempt to address?