Evaluation on measurement uncertainty of manual differential leukocyte counting

Hongwei SONG,Huoyan JI,Jianxin WANG,Huimin WANG
DOI: https://doi.org/10.3969/j.issn.1672-9455.2017.01.009
2017-01-01
Abstract:Objective To evaluate the measurement uncertainty of each type of manual differential leukocyte counting . Methods Three blood samples were collected and prepared three blood smears for each sample ,which were performed the differen‐tial leukocyte counting by three different experimenters .Considering the influence of the four factors sample adding volume ,differ‐ence among experimenters ,cellular distribution type and rounding off numerical values on the uncertainty of differential leukocyte counting ,each factor′s uncertainty of every type leukocyte in each blood sample was firstly calculated .Then the total uncertainty was synthesized by various components according to the uncertainty propagation law .Results The percentages of various types of leukocyte in three blood samples from high to low were neutrophilic granulocyte ,lymphocyte ,monocyte ,eosinophilic granulocyte and basophilic granulocyte .The synthesized uncertainties for each type leukocyte were respectively 7 .9% ,8 .8% ,5 .1% ,4 .3% and 3 .8% for the sample 1 ;7 .4% ,7 .4% ,5 .2% ,4 .5% and 3 .8% for the sample 2 ,and 7 .6% ,6 .4% ,5 .7% ,4 .3% and 3 .8% for the sample 3 .Conclusion Excepting lymphocyte in the sample 1 ,the synthesized uncertainty of each type leukocyte is proportional to the percentage in the total leukocyte counting .
What problem does this paper attempt to address?