Ultra-high Performance Liquid Chromatography Tandem Mass Spectrometry Method for Detection of Α-Hydroxybutyrate

Jin Zhonggan,Zong Ming,Fu Zhiyan,Ye Zhihan,Fan Lieying
DOI: https://doi.org/10.3760/cma.j.issn.1009-9158.2019.01.012
2019-01-01
Abstract:Objective To establish an ultra high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method for detecting α-hydroxybutyrate (α-HB) in serum.Methods Electrospray ionization negative ion and multiple reaction monitoring mode were used to detect serum α-HB.The linearity,low limits of quantification,precision,recovery and interference of UHPLC-MS/MS were evaluated.The reference interval of this method was established in 130 serum samples (62 males and 68 females) from Shanghai East Hospital.Dixon method was used to judge the outliers and K-S test was used to analyze the data normality.The standard curve was scored by linear regression analysis.Results The total run time was 4 min of UHPLC-MS/MS method for the determination of α-HB.It has a good linear relationship in the range of 0.5-40.0 mg/L(r=0.999 4);the low limit of quantification was 0.5 mg/L;the in-batch and inter-batch coefficient of variation precision were less than 4.1% and 6.3%,respectively;the recovery ranged between 95.8% and 103.8%.Hemolytic samples (about 5 g/L hemoglobin),lipemic samples (about 12 mmol/L triglyceride),icteric samples (about 150 μmol/L total bilirubin) had no significant interference to the detection.The reference range of the apparent healthy population was 1.46-6.48 mg/L.Conclusions A method for the determination of serum α-HB by UHPLC-MS/MS was established.The method was simple,rapid,and could be used for the detection of clinical samples.