Study of Artery Elasticity Changes of Rabbits' Abdominal Aorta after Biodegradable JDBM Stent Implanted by Echotracking Combined with Micro CT

Li CHEN,Jin-yun TAN,Feng YUAN,Yong WANG,Jia PEI,Bo YU,Guang-yin YUAN,Jun ZHANG,Yi WANG
DOI: https://doi.org/10.3969/j.issn.1006-5741.2017.02.017
2017-01-01
Abstract:Purpose:To determine whether echotracking combined with microCT could be used to assess the artery elasticity during the degradation processes.Methods:Either bare biodegradable magnesium alloy JDBM stent (JDBM group,n=12) or bare nitinol stent (BMS group,n=12) were implanted in New Zealand white rabbits' abdominal arteries.Echotracking was performed to obtain the stiffness parameter β before operation,immediate post-operation,14,30 and 60 days post-operation.The degradation of the stents was analyzed by mirco CT at the 14,30 and 60 days after operation.Results:① There was no migration or fracture of all the stents,and the blood flow in stent was normal.② MircoCT showed that the nitinol stents were with no changes.The JDBM stents began to degrade since 30 days after operation.But their structures maintained well and the struts were in continuity.③ There was no statistical significant difference of β between each groups before operation,immediate post-operation,or 14 days post-operation (P > 0.05).As the implanted time went by,β of JDBM group was significantly lower than that of BMS group(P =0.036) at 30 days post-operation.However,β of BMS group was immeasurable at the 60 days post-operation because of the undetectable vessel diameter changes.Conclusion:Echotracking combined with microCT could be used to assess the artery elasticity during the degradation processes of JDBM.