Abstract:【Objective】To evaluate the safety and efficacy of Perclose device during hemostasis after coronary arteriography and interventional procedures,comparing with manual compression.【Methods】One hundred seventy-two consecutive patients who underwent coronary arteriography and intervention were divided into Perclose device group(80 cases,coronary arteriography 47 cases and coronary intervention 33 cases) and manual compression group(92 cases,coronary arteriography 56 cases,coronary intervention36 cases).The time to achieve hemostasis,the immobile time and vascular complications and success rate of hemostasis were compared between the two groups.【Results】The time to achieve hemostasis[CAG:(4±1) vs(30±3) min;PCI:(4±1) vs(35±10) min,all P0.001],the immobile time[CAG:(5.1±0.6) vs(13.3±1.3) h;PCI:((6.5)±1.2) vs(18.3±2.2) h,all P0.001] and vascular complications(0 vs 16.3%) were significantly reduced in Perclose group comparing with manual group.Success rate of hemostasis was similar(97.25% vs 96.74%,P0.05).【Conclusion】Hemostasis with Perclose device is safe and efficient after coronary arteriography and interventional procedure.