Value of automated breast volume scanner and elastography in differential diagnosis of breast lesions

Chaoli Xu,Shuping Wei,Yingdong Xie,Ninghua Fu,Bin Yang
DOI: https://doi.org/10.16245/j.cnki.issn1008-6978.2015.01.004
2015-01-01
Abstract:Objective To investigate the inter-observer reliability of automated breast volume scanner(ABVS) and elastography interpretation in differential diagnosis of breast lesions,and compare the diagnostic value among ABVS, elastography and the combination of these two methods . Methods Forty-one patients received ABVS and elastography examination. Kappa statistics was used to interpret concordance between examiners with ABVS and elastography, and concordance between pathology and ABVS, elastography, and the combination of these two methods.χ2 test was used to compare diagnostic value among these three methods. Results Inter-observer reliability of examiners with ABVS (K=0.620) and elastography (K=0.654) were consistent. ABVS (K=0.810), elastography (K=0.765) and the combination of ABVS and elastography (K=0.901) were very consistent with pathology diagnosis. The diagnostic accuracy of ABVS, elastography,and the combination of ABVS and elastography were 91.30%,89.13%,95.65%, the sensitivity were 100%,96.42%,100%, the specificity were 77.78%,77.78%, 87.50%,the postive predictive value were 87.50%,87.10%,93.75%,and the negative predictive value were 100%,93.33%,100%, respectively. Conclusion The ABVS and elastography are reliable imaging methods for the differentiation of the breast lesions with high diagnostic value and a good inter-observer and inter-rater reliability. A combination of ABVS and elastography would be favorable to improve diagnostic accuracy.