Comparative Study of MRI Image Quality about Small Hole Phase Array Coil and Clinical Flexibility Coil in Small Animal

张安君,赵喜,王国年,龚志刚,詹松华
DOI: https://doi.org/10.3969/j.issn.1672-0512.2010.03.002
2010-01-01
Abstract:Objective:To compare the imaging quality of the small aperture orthogonal phased-array creatural experiment coil and one-channel surface flexibility coil.Methods: Compare the two kinds of coil in the rat's head scanning of the signal to noise ratio,image quality and spectral imaging of brain using the same scanning parameters with 1.5T magnetic resonance imaging system.Results: Imaging quality on small hole phase array coil was significantly superior to that on one-channel surface flexibility coil(P0.01).The SNR value of the brain tissue in rats of the rat small aperture orthogonal phased-array creatural experiment coil was better than that of the surface flexibility coil(P0.01).The image quality score of the small aperture orthogonal phased-array creatural experiment coil was better than that of the surface flexibility coil(P 0.01).Conclusion: The head images obtained by small aperture orthogonal phased-array creatural experiment coil is better than the images obtained by the surface flexibility coil.Small hole phased-array coil can be used well in clinical experiment.