The performance evaluation of FUS3000Plus automatic urine system

林金表,冯丹琴,蔡建国,叶辉铭
DOI: https://doi.org/10.3969/j.issn.1673-5013.2019.06.009
2019-01-01
Abstract:Objective: To evaluate the performance of FUS3000Plus automatic urine system which is produced by Changchun DIRUI company before itˊs being used in laboratories. Methods: According to the medicals standard of the Peopleˊs Republic of China General Technical Requirement and National Committee for Clinical Laboratory Standards ( NCCLs) , ten parameters of urine dry chemistry analysis were evaluated by FUS3000Plus automatic urine system. The evaluation included within-run precision, between-run precision, carry-over, stability and accuracy. The precision, categorical correlationand Microscopy coincidence rate between FUS3000Plus automatic urine system and manual microscopy of urinary sediment were evaluated. Results: 1. Urine dry chemistry analysis: within-run precision, between-run precision, carry-over, stability and accuracy of urinalysis could meet the requirement of professional standards; 2. Urinary sediment: Within-run precision, for the urine with the concentration being 50 and 200 per μl, the coefficient of variation (CV) of RBC was 5.3% and 4.3% respectively. There were 10.3% and 3.7% for WBC. Between-run precision, for the urine with the concentration being 50 and 200 per μl, the CV of RBC and WBC were 8.85%, 4.61%, 7.1% and 5.35% respectively. The correlation coefficient of RBC and WBC was 1 with 0-15000/μl and 0.9999 with 0-5000/μl respectively. Microscopy coincidence rate of RBC, WBC and CAST between FUS3000Plus automatic urine system and manual microscopy was 87.7%、 89.8%、 68%, and the false negative rate of them was 2.46%、 1.97%、 0.98%respectively. Conclusion: FUS3000Plus automatic urine system is reliable, which can meet the need of clinical use, and has great application value.
What problem does this paper attempt to address?