Comparison between Korotkoff sound method and vulnerarious method in the blood measurement in atrial fibrillation patients

Yangxin CHEN,Guosheng FU,Jianjun JIANG,Lijiang TANG,Chongfeng FANG
DOI: https://doi.org/10.3969/j.issn.1001-1439.2005.09.007
2005-01-01
Journal of Clinical Cardiology
Abstract:Objective:To investigate the difference between the two methods (Korotkoff sound method and vul nerarious method) in atrial fibrillation patients, and discuss the limitation of Korotkoff sound method. Method: Measuring the blood pressure with Korotkoff sound method and vulnerarious method in 82 patients with atrial fi brillation and 40 patients with sinus rhythm accepting coronary angiography, and separating the patients with atrial fibrillation into some groups according to the degree of pulse deficit. Result:The systolic blood pressure(SBP) and diastolic blood pressure(DBP) are both lower with Korotkoff sound method than with vulnerarious method ( P 0. 01 ), and the error are increased with the increasing of degree of pulse deficit. But the differences between the blood pressure level, in patients with sinus rhythm, measured by Korotkoff sound method and vulnerarious method have no statistical significance ( P 0. 05). Conclusion:It's not suitable to measure blood pressure with Korot koff sound method in atrial fibrillation patients,especially in the patients with higher heart rate. And the method should be adjusted to meet the demand of measuring blood pressure correctly.