The application of orbital implantation of left endocardial electrode for cardiac resynchronization therapy

Jie YANG,Zi-ping CHENG,Gang CHEN,Ren ZHAO,Min LIU,Jun FENG,Li DU,Xue-gong SHI,Bang-ning WANG
DOI: https://doi.org/10.13333/j.cnki.cjcpe.2019.01.005
2019-01-01
Abstract:Objective To summarize the method and effect of orbital implantation of cardiac resynchronization therapy (CRT) or CRT and cardioverter defibrillator (CRT-D) combined with implantation of left ventricular endometrium electrode in patients with heart failure (HF). Method Six patients with HF who were unable to perform implantation of CRT/CRT-D left ventricular epicardial electrode coronary venous sinus in our hospital were selected.Femoral vein-atrial septum-femoral artery orbit and femoral artery-atrial septum-subclavian vein orbit were established successively by puncturing atrial septum, and then complete left ventricular endocardial electrode implantation.Cardiac ultrasound, electrocardiogram and cardiac function were compared before and after surgery. Result All the 6 patients successfully implanted CRT/CRT-D left ventricular endocardial electrode by orbital method.QRS duration was significantly narrowed after 1 week follow-up[ (140.83±10.52) ms vs (166.83±13.60) ms, P<0.05], and the ejection fraction was significantly increased[ (0.37±0.05 vs (0.27±0.04), P<0.05].After surgery, one patient suffered from left ventricular electrode perforation.Bedside cardiac ultrasound showed a small amount of effusion in the pericardium, which was improved after resetting the left ventricular electrode. Conclusion Orbital method implantation of CRT/CRT-D left ventricular endocardial electrode is safe and effective.