Evaluation of thoracic pedicle screw placement position by computed tomography

毛克亚,王岩,张永刚,刘郑生,肖嵩华,刘保卫
DOI: https://doi.org/10.3969/j.issn.1004-406X.2005.04.008
2005-01-01
Abstract:Objective:To evaluate the accuracy of thoracic pedicle screw placement by computed tomography.Method:Thoracic pedicle screws were inserted in 37 patients subjected to thoracic pedicle screw fixation,and postoperative computed tomography(CT) were scanned to determine the screw position,cortical perforation number and distance.Result:A total of 405 thoracic pedicle screws were inserted,in which 124 screws(30.61%) violated the pedicle wall.76 screws(18.77%) violated the pedicles lateral wall,32 screws(7.90%) violated the medial wall,and 16 screws(3.95%) violated the vertebral anterior wall happened respectively.Furthermore,perforated distance of 66 screws(16.30%) were less than 2mm,37 screws(9.14%) ranged 2mm~4mm,and 21 screws(5.9%) were beyond 4mm.Conclusion:Although thoracic pedicle screw violated pedicle wall happened frequently,thoracic pedicle screw could be inserted safely and accurately without neurologic,vascular or visceral complications if strictly according to the screw placement technique.