Biomechanical advantages of dual over single iliac screws in lumbo-iliac fixation construct

于滨生,郑召民,庄新明,李泽民,吕维加
DOI: https://doi.org/10.3760/cma.j.issn.0253-2352.2010.06.011
2010-01-01
Abstract:Objective To evaluate the biomechanical effects of single and dual iliac screws on the stability of lumbo-iliac fixation construct. Methods L2-pelvic specimens were harvested from 7 fresh hu-man cadavers. Following the test of the intact state simulated by L3-L5pedicle screw fixation, the destabi-lization by total sacrectomy and L3-iliac screw-rod reconstructions with four placements of iliac screws were conducted as follows: Group A, single short iliac screw; Group B, single long iliac screw; Group C, dual short iliac screws (respectively placed in the upper and lower iliac columns); Group D, dual short iliac screws (all placed in the lower iliac column). The establishment and testing of iliac screw technique were performed in random order. Biomechanical testing was performed on a material testing machine under 800 N compression and 7 N·m torsion loading modes for construct stiffness evaluation. Results In compression, the constructs of Group A-D obtained the initial compression stiffness of 73%, 76%, 98% and 112%, respectively. Group A and B were obviously lower than the intact state, and Group C and D (P<0.05). No significant differences were demonstrated between Group A and B (P>0.05), however, the difference between Group C and D was significant (P<0.05). In torsion, Group A-D acquired 72%, 79%, 105% and 109% torsion stiffness of the ini-tial condition, respectively. Group A and B were markedly lower than the other three groups (P<0.05). No significant differences were detected not only between the Group A and B, but also among Group C, D, and the intact condition (P>0.05). Conclusion For the instability followed by total sacrectomy, single short and long iliac screws do not help in restoring the local stability to the intact condition. However, dual iliac screws provided much higher anti-compression and torsion resistance properties than using single iliac screw tech-niques, and the construct stability could compare to the initial state.
What problem does this paper attempt to address?