Rapid and effective sutureless anastomosis of femoral artery by magnetic compression anastomosis technique

刘仕琪,雷鹏,吕毅,高睿,关正,王善佩,史建华,刘艳奎
DOI: https://doi.org/10.3760/cma.j.issn.1001-4497.2015.01.008
2015-01-01
Abstract:Objective To investigate the magnetic pinned-ring device for non-suture vascular anastomosis.Methods Twelve adult mongrel dogs of either gender were randomly divided for femoral artery in situ end-to-end anastomosis,one side of the femoral artery was anastomosised with MCA,and the other side of the femoral artery was anastomosised by hand suturing in each dog.Operation time and stoma errhysis were recorded during operation.Patency and stoma stenosis was confirmed via color Doppler ultrasound scans and X-ray cholangiography at different time points as late as 6 weeks after surgery.Results The time required to perform the vascular anastomosis was significantly shorter for the magnetic device than hand sewing[(3.89 ± 1.16) min vs.(14.16 ± 3.72) min,P =0.000].There were 6 patients who happened stoma errhysis during femoral artery hand-sewing anastomosis,and needed to be repaired.Vascular X-ray angiography and color Doppler ultrasound found blood flow magnetic anastomosis group normal with 100% patency rate and there wasn't stoma stenosis,but one stoma occlusive and four stoma stenosis in the manual suture groups.Conclusion The magnetic pinned-ring device offers a simple,fast,reliable,and efficacious technique for non-suture artery anastomosis.Application of the MCA technique can effectively reduce the complication caused by operation technique,resulting in smooth operation and improved prognosis.