Diagnostic Value of Virtual Touch Imaging Quantification for BI-RADS 4 Breast Lesions

曹秋月,周玮珺,孔文韬,吴敏
DOI: https://doi.org/10.11735/j.issn.1671-170X.2018.03.B011
2018-01-01
Abstract:[Objective] To evaluate the application of virtual touch imaging quantification(VTIQ) in the diagnosis of BI-RADS 4 breast lesions.[Methods] One hundred and fifty patients with pathologically diagnosed breast lesions were enrolled in the study.All patients underwent ultrasonographic examination and the results were analyzed with VTIQ software.With pathologic results as the reference standard,the receiver-operating characteristic (ROC) curve with BI-RADS classification and with VTIQ correction were developed and the diagnostic value for breast lesions before and after correction was analyzed.[Results] Among 150 cases of breast lesions,there were 74 benign lesions and 76 malignant lesions.The VTIQ values of malignant lesions were significantly higher than those of benign lesions(P<0.001).The area under the ROC curve(AUC) of BIRADS classification was 0.807 and the AUC after correction was 0.884.Taking VTIQ of 4.09m/s as cut-off value,the sensitivity,specificity,accuracy,positive predictive value and negative predictive value of BI-RADS for diagnosis of malignant breast lesions were 81.58%(62/76),74.32% (55/74),78.0%(138/150),76.54%(62/81)and 79.71%(55/69),respectively.The sensitivity,specificity,accuracy,positive predictive value,negative predictive value of VTIQ correction were 90.79%(69/76),90.79%(69/74),92%(138/150),93.24%(69/74) and 90.79%(69/76),respectively.[Conclusion] Virtual touch imaging quantification technique can improve the diagnosis of BIRADS 4 classes breast lesions.