Clinical Research on Evaluating the Implant Stability by Resonance Frequency

吴小吉,林野,邱立新,李健慧,邸萍
DOI: https://doi.org/10.3969/j.issn.1007-3957.2005.02.007
2005-01-01
Abstract:Objective:To determine and evaluate the changes of implant stability in implant with sinus lift,the early exposure implants and standard placed implant in human using resonance frequency analysis(RFA).Methods::87implants in36patients were classified into3groups according to different clinical procedures,group A con-tained implants experienced the early exposure(EE),group B needed sinus lift with bone graft(SL)and the control group C(SE)had a traditional healing time advocated by Branemark.At the implant placement every implant was measured by RFA and the second measurement was done at second-stage or prosthodontical treat-ment.Results:The normal range of implant primary stability is54.92-80.20ISQ.Group A had the highest stability,and the difference was significant,the other two groups had no significant difference between each oth-er.The secondarystabilityrange ofRFAbefore prosthondics seems to be57.94-79.38ISQ.There was no differ-ence among the three groups.As for the sinus lift and early exposure,no statistical difference occurred,but there was some change of stability in the control group in their two measurements.Conclusion:Resonance Fre-quencyAnalysis ofthe implantprimarystabilityand osseointegrationis a reliable and effective assistant.The tech-nique of sinus graft with simultaneous implant and early implant exposure in indicated cases should be pre-dictable.
What problem does this paper attempt to address?