Fatigue stress of nickel-titanium rotary endodontic instruments under curve conditions: a finite element analysis

Xiao-bo CHEN,Lin YUE,Xue-jun GAO
DOI: https://doi.org/10.3760/cma.j.issn.1002-0098.2013.s1.005
2013-01-01
Abstract:Objective To investigate the curvature's impact on the fatigue stress in nickel-titanium files.Methods A three-dimensional digital model was built for new nickel-titanium file (HER0642,MicroMega,France) of size 30 taper 0.02 with software Solidworks.The fatigue stress in the model under a variety of bending conditions was calculated using the finite element analysis software ANSYS 11.0 Workbench.Bending parameters included the radius and degree of the curvature.Values and distribution of both normal stress and shear stress were calculated and compared.Results The highest normal stress under 60°curvature condition was higher(1274.20,980.48 MPa),the highest shear stress was also higher(38.16,42.48 MPa),but the highest nomal stress under 20° curvature condition was lower (533.37,493.56 MPa),the highest shear stress was also lower(16.81,14.26 MPa).The highest normal stress was concentrated on the blade surface where the curvature began.The highest shear stress was also observed at the same location.The values and distributions had a positive relationship with the curvature degree,but a negative relationship with the curve radius.The stress concentration span enlarged by the arc length.Conclusions The normal stress was concentrated on the blade surface where the curvature began,suggesting high incidence of fatigue fracture.
What problem does this paper attempt to address?