Efficacy and safety assessment of expiratory positive airway pressure rhinobyon for OSAHS therapy

Yaxuan LIU,Yangyang YING,Yan WANG,Shuang DOU,Yanzhong LI,Dedong MA
DOI: https://doi.org/10.6040/j.issn.1673-3770.0.2017.025
2017-01-01
Abstract:Objective Evidence shows that CPAP therapy relying mainly on end-expiratory positive airway pressure (EEPAP) with OSAHS.Therefore,we designed expiratory positive airway pressure (EPAP) rhinobyon and made the efficacy and safety assessment.Methods Six volunteers were enrolled to measure the end-tidal carbon dioxide pressure (PErCO2) while being treated by EPAP rhinobyon under different expiratory pressure.Fourteen OSAHS patients (AHI ≥15/h by PSG) recruited were divided into two groups randomly and treated with CPAP and EPAP rhinobyon for a week respectively.After a week of washing out,the patients were applied with exchanged therapeutic methods for another week.The PSG was performed at the end of each week of treatment with device-on.Results There was no significant difference of PEr CO2 with EPAP therapies under different expiratory pressure and without any therapy (F =1.162,P =0.34).Among the 14 OSAHS patients in both groups,AHI decreased,minimum SaO2 and mean SaO2 increased significantly after treatment (P < 0.001,respectively).There was no significant difference of the efficacy between EPAP rhinobyon and CPAP therapy (P =0.830,0672,0.649;P =0.473,0.636,0.667).Conclusion EPAP rhinobyon therapy was safe and reliable with significant efficacy for selected OSAHS patients.It could be promoted as the substitutive therapy for OSAHS in the future.