Correlation analysis between change of corneal biomechanical parameters and measurement of intraocular pressure after LASIK

隋丹丹,胡琦,吴琼,李雪,李庆军,杨帆,康杨,杜春宇
DOI: https://doi.org/10.13281/j.cnki.issn.1004-4469.2016.03.009
2016-01-01
Abstract:Objective To observe the relationship between the change of corneal biomechanical parameters and the measurement of intraocular pressure after LASIK.Design Prospective case series.Participants Eighty-one patients (162 eyes)with myopia were performed LASIK from Janurary to December in 2014 in the First Affiliated Hospital of Harbin Medical University.Methods The corneal hysteresis (CH),corneal resistance factor (CRF),Goldmann-correlated IOP value (IOPg) and corneal-compensated intraocular pressure (IOPcc) were examined with ocular response analyzer (ORA) before and after LASIK.The intraocular pressures were measure using non-contact tonometer (NCT) as well.Main Outcome Measures NCT,IOPcc,IOPg,CH,CRF before and after LASIK.Results Postoperative 3 months,CH,CRF,IOPcc,IOPg,and NCT measured values decreased significantly compared with those preoperatively.Thecomparison among IOPcc,IOPg and NCT postoperatively shows the significant distinction (P<0.05).There is a negative correlation between △CH,△CRF and △IOPcc,△IOPcc-IOPg and △IOPcc-△NCT.There is a positive correlation between △CH,△CRF and △ IOPg and △ NCT (P<0.05).Conclusion The corneal biomechanics and the measurement of intraocular pressure after LASIK were lower than preoperatively.There is a correlation between the decline of the measurement of intraocular pressure and the decrement of CH and CRF.The level of intraocular pressure drop may be affected by corneal biomechanical properties.