The application of low dose mammography in the needle-localized breast biopsy for impalpable calcified lesions

张静,阎瑾,江怡菲,王丽君,罗冉,李芳珍,汪登斌
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2015.11.006
2015-01-01
Abstract:Objective:The aim of this study was to assess the application of low radiation dose mammography in the needle-localized breast biopsy (NLBB)with impalpable calcified lesions.Methods:Twenty patients with impalpable but cal-cified lesions on mammography underwent needle-localized breast biopsy were included in this study and randomly divided into two groups (dose reduction group and STD group).For the dose reduction group,the milliampere seconds (mAs)were reduced by 50% manually during the procedure.Standard mode were applied in STD group performed under automatic pa-rameters.The actual radiographic parameters,kVp and mAs,for both groups were documented and compared for statistical purpose.The localization satisfaction (i.e.,the distance between the needle tip and lesion center less than 2cm),and the difference of mean AGD and ESD values in both groups were compared.Results:The localization satisfaction between the dose reduction group and STD group had no significant difference (P =0.686 >0.05 ).The mean AGD and ESD between two groups was significantly different (P =0.005<0.05).Conclusion:The low radiation dose mammography in needle-lo-calized breast biopsy (NLBB)for impalpable calcified lesions doesn't affect the localization accuracy,and it yet reduces the radiation dose effectively.