iCare rebound tonometer in acute attack of primary angle closure glaucoma

Yan Zhang,Hongying Cai,Yi Ma
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2016.02.005
2016-01-01
Abstract:Objective To observe the result of iCare rebound tonometer (RBT) using in acute attach eyes of primary angle closer glaucoma (PACG).Methods RBT,non contact tonometer (NCT)and goldmann applanation tonometer (GAT) were used to test the PACG acute attack eyes.The mean intraocular pressure (IOP) measurements were analyzed by paired t-test.Pearson correlation analysis was performed to assess the correlations (SPSS17.0 soft ware).Agreement between the tonometers was calculated using the Bland-Altman method (MedCalc soft ware).Results Forty-five PACG acute attack eyes of 38 patients were included,11 males,27 females; aged 42-72 (averaged 59.0±8.2).There was no statistical difference between the IOP tested by RBT and GAT (P =0.961,t =0.049).High positive correlation were found between RBT and GAT (r =0.963).Bland-Altman scatter plots showed reasonable inter-method agreement between RBT and GAT.40.0% (18/45) eyes failed to test the IOP by NCT.There was statistical difference between the IOP tested by NCT and GAT (P =0.035,t =-2.224).High positive correlation were found between NCT and GAT (r =0.963,r =0.905).Bland-Altman scatter plots showed reasonable inter-method agreement between NCT and GAT.Conclusions RBT is more reliable tonometer than NCT to test the IOP of PACG acute attack eyes,it has a good positive correlation and reasonable inter-method agreement with GAT.This tonometer is convenient and easy to learn,less trauma to the testing eyes.We suggest RBT to be used for acute attack eyes of PACG.