A study of IOP and ocular pulse amplitude with DCT in primary open-angle glaucoma and normal tension glaucoma patients

Yun Hongmin,Fu Pei,Yuan Jinsong,Zhang Bin,Li Xiaoxin
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2006.11.005
2006-01-01
Abstract:Objective To compare the differences of IOP and OPA measured with Dynamic ContourTonometer (DCT) and Goldmann Applanantion Tonometer (GAT) in primary open-angle glaucoma patients,normal tension glaucoma patients and healthy controls; Also to investigate the correlations between IOP mea-sured with DCT and GAT, heart rate (HR), systolic blood pressure (SBP) and diastolic blood pressure (DBP);Methods IOP (DCT and GAT), OPA, HR, SBP and DBP were measured among the subjects of 31 POAGpatients, 31 NTG patients and 31 normal controls. Results There are significant statistical differences of IOPmeasured using either DCT or GAT, OPA, SBP and DBP among POAG patients, NTG patients and healthycontrols except HR. OPA are (2.85 ± 1.15)mmHg in normal controls, (1.99 ± 1.04)mmHg in NTG patientsand (3.00 ± 1.07)mmHg in POAG patients respectively. There is no significant statistical difference of OPAbetween POAG patients and normal controls (P=0.568), whereas significant statistical difference of OPA be-tween POAG patients and NTG patients (P<0.001) is observed. Significant statistical difference of OPA be-tween NTG patients and normal controls (P<0.001) is also detected. Conclusions DCT is equivalent in accu-racy to GAT in the measurement of IOP in POAG patinets, NTG patients and normal controls; The differencesof OPA among POAG patients, NTG patients and normal controls are identified, which has potential indicationin the diagnosis and treatment of glaucoma.