Determination of Biapenem in Human Plasma by HPLC

Aidijie Ba-yasi,LIU Yi,YU Gang,ZHAO Li-bo,ZHANG Chun-yan,WANG Qian,LI Yu-zhen,FANG Yi
2009-01-01
Abstract:To establish an HPLC method for the determination of biapenem in human plasma. METHODS: Sample separation was conducted on Hypersil ODS chromatographic column; the mobile phase consisted of acetonitrile-0.1 mol·L~(-1) sodium acetate buffer solution (2 : 98,pH 4.38) at a flow rate of 1.0 mL·min~(-1). The detection wavelength was set at 300 nm;the column temperature was kept at 25℃; the sample size was 20 μL and the sample temperature was kept at 10℃. RESULTS: Biapenem showed a good linear relation in the range of 0.2~50 μg·mL~(-1), with a lowest detectable limit of 0.2 μg·mL~(-1). The average recovery rates of biapenem at 4 different concentrations (0.2, 0.5, 5, 50 μg·mL~(-1)) were 105.8%, 91.5%, 86.84% and 97.00%, respectively, and the average recovery rate of internal standard was 97.04%. Both the intraday and inter-day RSD of the 4 concentrations of biapenem were less than 15%, all met the criteria for the determination of biological samples. CONCLUSION: The HPLC method is simple, accurate, rapid, sensitive, specific and reproducible, and it is applicable for the pharmaeokinetic study of biapenem in human plasma.