Study on Determination of ZLR-8 by HPLC

YANG Jing,LIU Wen-ying,ZHANG Yi-hua,HUANG Hai-yan,LI Qiang
DOI: https://doi.org/10.3321/j.issn:1001-2494.2006.12.022
2006-01-01
Abstract:OBJECTIVE To develop a RP-HPLC method for the quality control of(ZLR-8).METHODS The separation was performed on a Lichrospher ODS column(4.6 mm×250 mm).The mobile phase consisted of methanol-acetonitrile-water(55∶20∶10).The flow rate was 1.0 mL·min~(-1).The UV detection was set at 226 nm.RESULTS The intermediates and compounds produced by degradation did not interfere ZLR-8.A good linearity was found over the concentration range of 11.18 to 223.6 mg·L~(-1),the RSD for the repeatability was 0.47%(n=6),the limit of detection for ZLR-8 was 54 μg·L~(-1).CONCLUSION The method is simple and accurate.