Establishment of a 2D-LC-UV Method for the Determination of Tigecycline in Human Serum and Its Clinical Application

Xue-mei LUO,Lu JIN,Pei LIANG,Feng WANG,Wei-hong GE
DOI: https://doi.org/10.13664/j.cnki.pcr.2016.04.009
2016-01-01
Abstract:Objective: In this study, tigecycline, a new glycylcycline antimicrobial agent was quantified by a two-dimensional liquid chromatography coupled with ultraviolet detection (2D-LC-UV). Methods: In this setup, two reversed phase liquid chromatography columns, ASTON C8 (100 mm×4.6 mm, 5μm) and Agi-lent Extent-C18 (150 mm×4.6 mm, 5μm), were used as the first and second separation dimensions, respectively. An ASTON SCX (20 mm×4.6 mm, 5μm) column was used as the connector to transfer the eluent from the first dimensional column to the second one for further separation. The mobile phases of the two dimensions consisted of 30 mmol·L-1-acetonitrile-methanol NaH2PO4 (40∶30∶30, v/v, pH=5.4) and acetonitrile-30 mmol·L-1 NaH2PO4 (75∶25, v/v, pH=3.5), respectively. The flow rates were both 1 mL·min-1 and the detection wave-length was 244 nm. Serum peak, middle and trough concentration samples from 40 patients with intra-ab-dominal infections administrated with tigecycline (50 mg q12h) were analyzed and values of area under curve (AUC) were estimated. Results: The calibration curves were linear over the range of 0.03-30μg·mL-1. The average serum peak, middle and trough concentrations were (1.60±0.77)μg·mL-1, (0.84±0.36)μg·mL-1 and (0.57±0.25)μg·mL-1, respectively. And the estimated AUC values were (22.66±9.77)μg·h·mL-1. Con-clusions: This method can be used effectively in clinic and research fields.
What problem does this paper attempt to address?