Study on the absorbability of different infusion sets to sulfotanshinone ⅡA sodium injection

WenJuan SHI,ZhongDong LI,XiuYan TAN,WeiWei QIN,Bin WANG,MingKang ZHONG
DOI: https://doi.org/10.5428/pcar20170215
2017-01-01
Pharmaceutical Care and Research
Abstract:Objective: To investigate the absorbability of the infusion sets produced by 5 different manufacturers to sulfotanshinone ⅡA sodium injection, so as to provide reference for the choice of infusion sets.Methods:The solution of sulfotanshinone ⅡA sodium injection was prepared as the actual concentration used clinically, and flowed through the infusion sets at the flow rates of 30 and 60 drops per minutes.The output solutions of sulfotanshinone ⅡA at different time points were collected for concentration detection by HPLC method, and then the obtained detection results were compared with the solution before passing through the infusion sets.Results: There were no significant differences in the concentrations of sulfotanshinone ⅡA sodium injections both before and after passing through the infusion sets, either at the identical or different drop rates.Conclusion: The infusion sets made by 5 different manufacturers did not display any significant absorbability to sulfotanshinone ⅡA sodium injections.
What problem does this paper attempt to address?