Study on efficacy and safety of ShengTang insulin pump infusion device

WU Jin-xiao,LV Xiao-feng,ZHU Lv-yun
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2013.09.005
2013-01-01
Abstract:Objective:To assess the clinical safety and effectiveness of ShengTang disposable insulin pump infusion device.Methods:A prospective,multi-center,randomized clinical validation test were conducted.Two centers enrolled a total of 126 cases,including 69 cases of the test group,used ShengTang-A type of insulin pump with ShengTang disposable insulin pump infusion device,and 57 cases of the control group,used the imported insulin pumps with imported disposable insulin pump infusion device.Results:Effectiveness:patency(97.01% of the test group,98.18% of the control group),the looper subcutaneous puncture success rate(95.65% of the test group,100%of the control group),tightness of the product of eachparts(98.53% of the test group,100% of the control group) are no statistically significant difference compared to the control group;Patency and tightness the division is split are the same as the control group,all of 100%.Security:none of the test group and the control group patients experienced adverse events,the test group and the control group had no significant difference from getting started to seven days after the end.Conclusion:The domestic ShengTang disposable insulin pump infusion device on the efficacy and safety is not inferior to imported products.
What problem does this paper attempt to address?