Abstract:Objective To establish a rapid, simple and accurate method to determine the content of semi-finished product in iohexol injection. Methods Iohexol injection was determined by the refractive index method. The linear equation was established and the precision and recovery of the method were determined. The difference between refractive index method and potentionmetric titration method in measuring the iohexol injection content was compared. Results The linear equation was n = 0.1674 C + 1.3318. The linearity was at 120-420 mg·m L-1 (r= 0.9992). The average recovery was 99.8%-100.4% (n = 3), and the precision RSD was 0.010% (n = 6). No disturbance from the excipients was found. The relative deviation of the refractive index method and potentionmetric titration method was less than 1.6%. Conclusion The method is rapid and accurate for the determination of semifinished product in iohexol injection, which can be used for the in-process quality control.