Experimental Study of C-peptide Piezoelectric Immunosensor

张波,府伟灵,毛琼国,汤万里,俞凡
DOI: https://doi.org/10.3321/j.issn:1000-5404.2002.12.020
2002-01-01
Abstract:Objective To construct a new type of C peptide piezoelectric immunosensor to determine human C peptide Methods The sensor pool consisted of a plastic cycle and 10 MHz quartz AT cut crystal with gold coated electrodes on it The immobilization of the monoclonal antibody of C peptide onto the gold electrode surface of the crystal was accomplished via thiolated method The immunosensor were used in clinical laboratory Results Experimental results showed that the piezoelectric immunosensor had good response to C peptide with frequency shifts linearly dependent on C peptide concentration in a range of 0 375 12 000 ng/ml Other antigens such as insulin and proinsulin did not intervene with the response of the sensor The analytical results by this approach were in satisfactory consistence with those from radiometric immunoassay ( P 0 05),in a correlation coefficient of 0 94 Conclusion Piezoelectric immunosensor for determining C peptide is a small in size, low in cost, simple and rapid in operation device, in addition, of high sensitivity,high specificity,and unnecessary labeling for real time detection Thus, it is suitable for detecting serum C peptide in clinical laboratory