Monitoring of Permanent Vascular Access Function in Hemodialysis

庞慧华,钱家麒,张庆怡,黄佩文,张纪蔚,张皓
DOI: https://doi.org/10.3969/j.issn.1674-8115.2000.04.006
2000-01-01
Abstract:Objective To compare the different monitorings of access function in hemodialysis. Methods Examined the access function of 26 regularly herncdialysis patients, dynamic venous pressure(VDP),percentage urea access recirculation(R), doppler ultrasound were evaluated simultaneously in Oct. 1997 to Mar.1998. Results 5 stenosis or occlusive events were found. All had abnormalities in ultrasound image, 1 had elevation in VDP, R were normal and access flow were over 400ml/min. Conclusion VDP and R can be used in monitoring access function, but there is some limitation in detecting access dysfunction. Doppler ultrasound can find abnormal event directly, it is an effective method for access function monitoring.
What problem does this paper attempt to address?