Numerical simulation of the effect of driving signals by phased transducer on transcranial focal region

Xin DING,Zhen-zhen HAN,Xi-qi JIAN
2015-01-01
Abstract:Objective:To investigate the effect of driving signals by phased transducer arrays on the transcranial focal region and to solve the problem of transcranial defocusing and thermal damage to the skull and tissue around it. Methods:Method was put forward by modulating phase and amplitude of driving signals based on the time reversal and cross-correlation among signals. A numerical simulation model was established based on skull CT data and phased transducer. The finite difference time domain method was used to simulate the distribution of transcranial HIFU pressure and temperature fields to study the influence of driving signals by phased array on transcranial HIFU focal region. Results:Results of numerical simulation indicated that focus was formed accurately at the target by modulating the phase of driving signals of transducer arrays;the maximum pressure and temperature at focus were increased by modulating the amplitude of driving signals by transducer arrays;temperature in skull was decreased by modulating the amplitude of different arrays of transducer. Conclusion:By modulating driving signals of phased array, we can achieve precise and effective transcranial focusing and the acoustic energy deposited at focus would increase. By modulating driving signals of different arrays, temperature decreases in skull thus avoiding thermal damage to skull and the surrounding tissue.
What problem does this paper attempt to address?