Abstract:Glucosamine in glycosylated whey protein was released hydrochloric acid hydrolysis.The RP-HPLC separation was performed on an Elite C18 (Hypersil ODS25 μm 4.6 mm x 250 mm) utilizing an isocratic elution of acetic acid sodium acetate buffer solution (containing 68% 0.025 mol· L-1 pH 3.6) and methyl alcohol (containing 32%) as the mobile phase at a flow rated of 1 mL· min-1.Injiction volume and column temperature were set at 10 μL and 35 ℃.The results indicate that the linear ranage was from 0.1 to 100 μg· mL-1 for glucosamine and correlation coefficient (R2 was greater than 0.99).The average recoveries spiked at the three concentration levels of 0.30,0.50,1.00 mg ranaged between 88.05% and 110.79% with the relative standard deviations from 1.22% to 5.92%.The limit of detection was 0.02 μg·mL-1.Content of glucosamine in glycosylated whey protein was 2.03 mg·g1.Therefore,this method had the characteristics of simple operation,high reproducibility and sensitivity,It could be widely applied to determine glucosamine in glycosylated protein.